《中国期刊全文数据库》收录期刊
《中国核心期刊(遴选)数据库》收录期刊
《中文科技期刊数据库》收录期刊
医疗器械园地

数字减影血管造影球管工作原理及质量维护

  • 沈彧枫 ,
  • 曹少平
展开
  • 上海中医药大学附属岳阳医院医学装备处, 上海 200437
沈彧枫(1985—),女,学士,助理工程师,从事医学装备管理工作

收稿日期: 2024-05-22

  网络出版日期: 2024-10-17

基金资助

上海市临床重点专科2019年评审项目(shslczdzk04601); 上海交通大学中国医院发展研究院医院后勤管理研究所2024年度项目(CHDI-2024-C-14)

Digital subtraction angiography bulb tube working principle and quality maintenance

  • SHEN Yufeng ,
  • CAO Shaoping
Expand
  • Department of Medical Equipment, Yueyang Hospital Affiliated to Shanghai University of Traditional Chinese Medicine, Shanghai 200437, China

Received date: 2024-05-22

  Online published: 2024-10-17

摘要

基于数字减影血管造影(DSA)球管工作原理,通过分析工作实践中的3个典型DSA球管故障,查找出故障主要成因,并概括了DSA球管的常见故障类型;总结出在安装新球管时的注意事项,如需仔细检查油泵接线,并对球管进行必要的训练和预热,以确保其稳定运行;最后详细描述了DSA球管各部位的维护保养及验收流程和标准。

本文引用格式

沈彧枫 , 曹少平 . 数字减影血管造影球管工作原理及质量维护[J]. 外科研究与新技术(中英文), 2024 , 13(3) : 249 -253 . DOI: 10.3969/j.issn.2095-378X.2024.03.016

Abstract

After the introduction of the working principle of digital subtraction angiography (DSA) bulb tube, through the analysis of three typical faults of DSA bulb tube, the main causes of the faults were found, and the common fault types of DSA bulb tube were summarized. It was concluded that when installing a new bulb tube, it was necessary to check the wiring of the oil pump carefully, and carry out the necessary training and preheating of the bulb tube to ensure its stable operation. Finally the maintenance and acceptance procedures and standards of the various parts of DSA bulb tube were described in detail.

参考文献

[1] 王颖颖,张伟成.数字减影血管造影技术[J]. 中国医疗器械信息,2004,10(6):1-4.
[2] 徐贤叁,邵健.Allura Xper FD20型数字血管减影系统故障案例分析[J]. 中国医学装备,2018,15(9):149-150.
[3] 顾军民. 数字平板DSA的工作原理和开机后设备报错故障的分析及维修心得[J]. 中国医疗设备,2019,34(6):168-169.
[4] 魏本正,杨涛.CTX线管的故障判断及更换[J]. 医学影像学杂志,2007(8):772-772.
[5] 赵德坤,万金鑫,张毅,等.Philip Aura螺旋CT旋阳模块故障1例[J]. 医疗卫生装备,2011,32(02):134.
[6] 许美珍,朱栋梁,罗来树,等.新型冠状病毒肺炎患者介入诊疗手术工作流程与感染防控专家共识(第一版)[J]. 中国医疗设备,2020,35(6):1-6.
[7] 邓其辉. 西门子Artis Q floor数字减影血管造影系统球管故障的检修[J]. 医疗装备,2022,35(19):104-106.
文章导航

/