Objective To understand the causes of high voltage faults and clarify the idea of equipment failure maintanence through introducing repair cases of new UIH uCT528 high voltage (generator system), which is conductive to solving users’ problems while reducing maintenance costs. Methods The study achieved understanding of the working principle and interrelationships of uCT528 high voltage by analyzing typical circuit diagrams of repair cases, identifying faulty components, and replacing them. Results Through chip-level maintenance, equipment failures were resolved, maintenance costs were reduced, and maintenance efficiency was improved. Conclusion This typical maintenance case study can provide reference for medical colleagues in dealing with high voltage failures, such as accurately determining high voltage failures, understanding measurement points and circuit diagrams, and clarifying ideas to accurately find faults and repair them.
[1] 沈彧枫,曹少平. 数字减影血管造影球管工作原理及质量维护[J]. 外科研究与新技术(中英文),2024,13(3):249-253.
[2] 魏本正,杨涛. CT机X线管的故障判断及更换[J]. 医学影像学杂志,2007(8):772-772.
[3] 邓其辉. 西门子Artis Q floor数字减影血管造影系统球管故障的检修[J]. 医疗装备,2022,35(19):104-106.
[4] 张富治,盛兴,王瑞海,等. X射线CT球管的研制进展[J]. 真空电子技术,2016(1):7-10.
[5] 王晓光. HD宝石能谱CT典型故障维修2例[J]. 医疗卫生装备,2017,9(9):153-154.
[6] 尹吉生. 扫描系统故障维修一例[J]. 中国医疗设备,2017,32(12):112-114.
[7] 吴康鑫. 多排螺旋CT机高压故障的原因与维修[J]. 世界最新医学信息文摘,2017,55(55):155-155.
[8] 李爱军. 预防性维护在医院医疗设备管理中的重要作用[J].中国医疗器械信息,2018,24(6):145-147.