Surgical Research and New Technique ›› 2019, Vol. 8 ›› Issue (3): 177-179.doi: 10.3969/j.issn.2095-378X.2019.03.009

• Original article • Previous Articles     Next Articles

Comparison of two methods to determine the position of left double-lumen endobronchial tubes

SHUAI Bing, KE Yuwen, ZHU Caiyan, ZHAO Zhao, LIU Zhiheng   

  1. Department of Anesthesiology, Second People’s Hospital of Shenzhen, Shenzhen 518035, China
  • Received:2019-04-30 Online:2019-09-28 Published:2019-12-05

Abstract: Objective To compare the positioning accuracy of left double-lumen endobronchial tubes by bronchial cuff pressure method and fiberoptic bronchoscopy.Methods A total of 100 patients with single lung ventilation surgery were randomly divided into two groups, with 50 cases in each group.The position of double-lumen endobronchial tubes was assessed by bronchial cuff pressure method and fiberoptic bronchoscopy respectively after double-lumen bronchial intubation.Results The positioning accuracy rate of fiberoptic bronchoscopy (100%) was significantly higher than that of bronchial cuff pressure method (88%, P<0.05).There was no significant difference between the two groups in complications after intubation (P>0.05).Conclusion The bronchial cuff pressure method is simple and practical, with high positioning accuracy and low complications.

Key words: Fiberoptic bronchoscopy, Double-lumen bronchus, Positioning